拔出内窥镜后,立即用纱布蘸上洗涤液或多酶洗液擦拭插入部,拔下送气、送水按钮并放入洗涤液中,插入清洗专用按钮,来回送气送水,按下吸引按钮座,来回吸引洗涤液与空气。防止黏液干燥后难以清除,同时点检内窥镜的外观。
After pulling out the endoscope, immediately wipe the insertion part with gauze dipped in detergent or multi-enzyme detergent, pull out the air and water supply buttons and put them into the detergent, insert the special button for cleaning, send air and water back and forth, press the suction button seat to attract the detergent and air back and forth. Prevent the mucus from being difficult to remove after drying, and check the appearance of the endoscope at the same time.
内窥镜维修公司提醒使用者,内窥镜在入水前一定要戴上防水帽,并认真检查防水帽是否盖紧。防水帽要时刻确保处于干燥状态,避免引起电气接口潮湿使EL座接点腐蚀。手工清洗时要确认清洗工具没有脏污,没有磨损的划痕、裂缝、扭曲、金属与刷毛脱落等情况,避免对内窥镜造成破损,清洁刷应保持平直,避免与按钮安装座产生摩擦。
The endoscope maintenance company reminds the user that the endoscope must wear a waterproof cap before entering the water, and carefully check whether the waterproof cap is tight. The waterproof cap shall be kept in a dry state at all times to avoid corrosion of the EL base contact caused by damp electrical interface. During manual cleaning, make sure that the cleaning tool is free of dirt, worn scratches, cracks, distortion, metal and brush hair falling off and other conditions to avoid damage to the endoscope. The cleaning brush should be kept straight to avoid friction with the button mounting base.
内窥镜的注意事项
Precautions for endoscope
1、在清洗和消毒的时候,建议把镜子和别的手术器械分开,用塑料盆清洗。
1. When cleaning and disinfecting, it is recommended to separate the mirror from other surgical instruments and use a plastic basin for cleaning.
2、消毒前要把镜头清洗干净,否则异物在消毒后积累在镜头上,以至影响图像质量。当镜头上异物过多时,可用棉签涂一些镜子附带的清洁剂在镜面上,然后用清水清洗干净。
2. Clean the lens before disinfection, otherwise foreign matter will accumulate on the lens after disinfection, which will affect the image quality. When there are too many foreign matters on the lens, you can apply some detergent attached to the mirror with a cotton swab on the mirror surface, and then clean it with clean water.
3、尽量不要频繁更换消毒灭菌方法,这样对镜子的密封有好处。
3. Try not to change the sterilization method frequently, which is good for the sealing of the mirror.
4、禁止用超声波清洗镜子,尽量不要用生理盐水清洗镜子,否则容易生锈迹。
4. It is forbidden to use ultrasound to clean the mirror, and try not to use physiological saline to clean the mirror, otherwise it is easy to rust.
5、当高温高压消毒完镜子后,一定要让镜子自然冷却,禁止用冷水冷却。
5. When the mirror is sterilized by high temperature and high pressure, it must be allowed to cool naturally. It is forbidden to use cold water for cooling.
6、所以的镜子浸泡消毒时间不能超过60分钟。
6. Therefore, the immersion disinfection time of the mirror should not exceed 60 minutes.
当天检查结束后应进行漏水检测,在水中测漏时不可安装或拆除测漏器连接接口,避免内窥镜进水,造成损坏,测漏结束后应充分排除镜中气体,避免弯曲外皮损坏。检测完毕,须用压缩空气将活检孔、吸引孔、送气送水等管腔吹干,特别是电气接点上的水分,然后垂直悬挂在干燥、温度适中的地方,否则残留的水分会使空气中的细菌在内窥镜内外繁殖,造成污染。更多相关内容就来我们网站
http://www.jiangrenyiliao.cn进行咨询吧!
After the inspection on the same day, water leakage detection shall be carried out. During the leakage detection in water, the connection interface of the leak detector cannot be installed or removed to avoid water ingress into the endoscope and damage. After the leakage detection, the gas in the mirror shall be fully discharged to avoid bending skin damage. After the detection, compressed air must be used to blow the biopsy hole, suction hole, air supply and water supply and other lumens dry, especially the moisture on the electrical contact, and then hang it vertically in a dry and moderate temperature place, otherwise the residual moisture will cause the bacteria in the air to breed inside and outside the endoscope, causing pollution. Come to our website for more relevant content http://www.jiangrenyiliao.cn Consult!