软式内镜使用后应当立即用湿纱布擦去外表面污物,并反复送气与送水至少10秒钟;再用吸引通道反复吸清水或清洗液,直到流人吸引瓶的水干净为止;后取下内镜,拆下各种附件,并装好防水盖,置合适的容器中送清洗消毒室。下面
内窥镜设备维修人员为您讲解软式内镜的水洗以及酶洗方式。
After using the flexible endoscope, immediately wipe the dirt on the external surface with wet gauze, and repeatedly send air and water for at least 10 seconds; Then repeatedly suck clean water or detergent through the suction channel until the water flowing into the suction bottle is clean; After that, take down the endoscope, remove various accessories, install the waterproof cover, and put it in a suitable container and send it to the cleaning and disinfection room. The endoscope equipment maintenance personnel will explain the water washing and enzyme washing methods of the flexible endoscope for you.
水洗
washing
(1)在流动水下彻底冲洗,用纱布反复擦洗镜身,同时将操作部清洗干净;
(1) Rinse thoroughly under flowing water, scrub the mirror body repeatedly with gauze, and clean the operating part at the same time;
(2)取下活检入口阀门、吸引器按钮和送气送水按钮,用清洁毛刷彻底刷洗活检孔道和导光软管的吸引器管道,刷洗时必须两头见刷头,并洗净刷头上的污物;
(2) Remove the biopsy inlet valve, aspirator button and air and water supply button, and thoroughly brush the aspirator pipe of the biopsy hole and light guide hose with a cleaning brush. When brushing, the brush head must be seen at both ends, and the dirt on the brush head must be cleaned;
(3)安装全管道灌流器、管道插塞、防水帽和吸引器,用吸引器反复抽吸活检孔道;
(3) Install the full-pipe perfusion device, pipe plug, waterproof cap and suction device, and use the suction device to repeatedly aspirate the biopsy hole;
(4)全管道灌流器接50毫升注射器,吸清水注入送气送水管道;
(4) The whole pipeline perfusion device is connected with a 50ml syringe, and the water is sucked into the air supply and water supply pipeline;
(5)用吸引器吸干活检孔道的水分并擦干镜身。
(5) Use an aspirator to absorb water from the biopsy hole and dry the lens body.
将取下的吸引器按钮、送水送气按钮和活检入口阀用清水冲洗干净并擦干。
Rinse and dry the removed aspirator button, water and air supply button and biopsy inlet valve with clean water.
内镜附件如活检钳、细胞刷、切开刀、导丝、碎石器、网篮、造影导管、异物钳等使用后,先放入清水中,用小刷刷洗钳瓣内面和关节处,清洗后擦干。
After the use of endoscopic accessories such as biopsy forceps, cell brushes, cutting knives, guide wires, lithotripters, net baskets, contrast catheters, foreign body forceps, etc., first put them into clean water, wash the inner surface of the clamp valve and joints with a small brush, and then wipe them dry.
清洗纱布应当采用一次性使用的方式,清洗刷应当一用一消毒。
The cleaning gauze should be disposable, and the cleaning brush should be disinfected once in use.
酶洗
Enzymolysis
(1)多酶洗液的配置和浸泡时间按照产品说明书。
(1) The configuration and soaking time of multi-enzyme lotion shall be in accordance with the product manual.
(2)将擦干后的内镜置于酶洗槽中,用注射器抽吸多酶洗液100毫升,冲洗送气送水管道,用吸引器将含酶洗液吸入活检孔道,操作部多酶洗液擦拭。
(2) Place the dried endoscope in the enzyme washing tank, use a syringe to aspirate 100ml of multi-enzyme washing solution, flush the air supply and water supply pipeline, use an aspirator to inhale the enzyme-containing washing solution into the biopsy hole, and wipe the multi-enzyme washing solution of the operation department.
(3)擦干后的附件、各类按钮和阀门用多酶洗液浸泡,附件还需在超声清洗器内清洗(5~10)分钟。
(3) The dried accessories, various buttons and valves shall be soaked in multi-enzyme detergent, and the accessories shall be cleaned in the ultrasonic cleaner for (5~10) minutes.
(4)多酶洗液应当每清洗1条内镜后更换。
(4) The multi-enzyme lotion should be replaced after cleaning one endoscope.
That's all about the water washing and enzyme washing methods of flexible endoscope. If you have any doubts or needs about this, please come to our website http://www.jiangrenyiliao.cn consulting service