软性内窥镜测漏时漏气的故障分析

来源:http://www.jiangrenyiliao.cn/ 发布时间:2022-11-26 浏览量:0

软性内窥镜工作方式是将插入软管插入人体管腔内,通过导光束传递光源照明,并通过导像束或CCD成像。插入管内有器械通道,配合器械使用达到或观察病灶的目的。下面内窥镜设备维修人员来讲讲软性内窥镜测漏时漏气的故障。
The working mode of the flexible endoscope is to insert the insertion hose into the human lumen, transmit light source illumination through the guide beam, and image through the guide beam or CCD. There is an instrument channel in the insertion tube, which can be used together with the instrument to treat or observe the lesions. Next, the endoscope equipment maintenance personnel will talk about the failure of air leakage during the leak detection of the flexible endoscope.
内镜插入部故障有很多比例是来自于器械管道故障,使用者在使用器械通路时,如果器械与内镜的角度配合不好或者器械回收拉拽的过程出现错误,很容易损坏器械管道,从而造成泄漏。
Many of the failures of the endoscope insertion part are caused by the failure of the instrument pipeline. When users use the instrument channel, if the angle between the instrument and the endoscope is not well matched or there is an error in the process of retrieving and pulling the instrument, it is easy to damage the instrument pipeline, resulting in leakage.
内窥镜设备维修
方法:
resolvent:
使用者在放置器械过程在如果受阻,应立刻停止用力,强行通过很容易刺破或划伤器械管道壁;器械在回收时应当处于闭合状态,如果器械张开回收,器械管道就会划伤或划破。
If the user is obstructed in the process of placing the instrument, he/she should stop using force immediately, and it is easy to puncture or scratch the pipe wall of the instrument by force; The instrument should be closed when it is recycled. If the instrument is opened for recycling, the instrument pipe will be scratched or broken.
在确定没有图像故障的情况下,此类故障属于C级维修。此类故障第三方维修,因为原厂维修需更换整个插入部的所有组件,而第三方维修只需更换钳子管道和弯曲橡皮就能故障,而且会便宜很多。
If it is determined that there is no image fault, such fault belongs to Level C maintenance. Third party maintenance is recommended for this kind of fault, because the original factory maintenance needs to replace all components of the entire insertion part, while the third party maintenance can solve the fault only by replacing the pliers pipe and bending rubber, and the price will be much cheaper.
此外软性内窥镜在医疗领域的应用越来越广泛,很多科室如内镜室、外科、耳鼻喉科等等都在使用软性内窥镜进行检查、手术,因此相应的软性内窥镜的维护和保养也越来越重要。所以定期对其进行清洗维护非常关键,更多相关内容就来我们网站http://www.jiangrenyiliao.cn咨询吧!
In addition, soft endoscopes are increasingly used in the medical field. Many departments, such as endoscopy, surgery, otorhinolaryngology, are using soft endoscopes for inspection and surgery. Therefore, the maintenance and care of corresponding soft endoscopes are becoming increasingly important. Therefore, it is critical to clean and maintain them regularly. More relevant information can be found on our website http://www.jiangrenyiliao.cn Consult!
s
新闻资讯
联系我们
  • 公司公司:匠仁医疗设备有限公司
  • 匠仁医疗山东济南维修部
  • 联系人:樊经理
  • 电话:13153199508
  • 地址:山东省济南市槐荫区美里东路3000号德迈国际信息产业园6号楼101-2室
  • 匠仁医疗湖南长沙维修部
  • 联系人:李经理
  • 电话:13873135765
  • 地址:湖南省长沙市雨花区劳动东路820号恒大绿洲14栋2409室
新闻资讯

樊经理:13153199508     李经理:13873135765

公司地址:山东省济南市槐荫区美里东路3000号德迈国际信息产业园6号楼101-2室  湖南省长沙市雨花区劳动东路820号恒大绿洲14栋2409室

     

鲁公网安备 37010402441281号