由于设备使用的特殊性,内镜很容易在使用过程中出现各种故障,但是这些故障一般有经验的人员都是可以自己的,下面
内窥镜设备维修厂家就将为大家介绍几种比较特殊的故障情况,希望能有所帮助。
Due to the particularity of the equipment, it is very easy for the endoscope to have various failures in the use process, but these failures can be solved by experienced personnel. The endoscope equipment maintenance manufacturer will introduce several special failures for you, hoping to help.
某些型号的结肠镜有时维修后初次使用便会出现无水或是少水的现象。而且在很短的时间内视野变得模糊,甚可以看见水,外观上可以出现弯曲橡皮肿胀。
Some types of colonoscope sometimes have no water or little water after first use after maintenance. In addition, in a very short period of time, the vision becomes blurred, and even water can be seen, and the appearance can show bending rubber swelling.
此时用测漏器也无法判断出什么部位出现问题,打开此镜会出现镜体内已注满了水,如果处理不及时,显微镜内像束大多会产生龟裂,螺线管铜丝出现生锈等情况,严重时会造成报废。
At this time, the leak detector can not determine what part is wrong. When the mirror is opened, the mirror body will be filled with water. If it is not handled in time, most of the image bundles in the microscope will crack, and the copper wire of the solenoid will rust. If it is serious, it will be scrapped.
故障原因
Cause of failure
的某些结肠镜在操作把下方有一附送水部位,这个部位在修理安装的过程中,可能会发生变形或有硬的小杂物,装配时产生裂痕,水便从此处流入镜体。所以送气,送水时会出现水少或无水的情况,而且储水到一定量时,水压会使橡皮出现膨胀。
Some colonoscopes in the operation handle have an attached water supply part. During the repair and installation process, this part may be deformed or have hard small debris, which may cause cracks during assembly, and water will flow into the mirror body from here. Therefore, when air and water are delivered, there will be little or no water, and when water is stored to a certain amount, the water pressure will cause the rubber to expand.
防止故障方法
Failure prevention method
修理人员在维修完毕后或使用人员在前都应该进行测漏,特别是结肠镜,如果在测漏的过程中发现喷嘴水汽管出有气泡冒出而管道完好的情况下,应考虑附送水部位的接口有问题,应重新检查后安装,否则待水进入镜体后侧漏器也无法测出。
After the repair is completed or before the use, the repair personnel should conduct leak detection, especially for colonoscope. If bubbles are found in the nozzle steam pipe during leak detection and the pipe is in good condition, it should first consider that there is a problem with the interface of the attached water supply part, which should be rechecked before installation, otherwise the leak can not be detected when the water enters the back of the mirror body.
There are many reasons for equipment failure, which requires users to strictly follow the operating specifications when using the equipment, so as to reduce the number of equipment failures as much as possible. More relevant content will come to our website http://www.jiangrenyiliao.cn consulting service