医疗用内窥镜设备定期保养一般由仪管员配合工程技术人员完成,它是一项不断循环进行的有组织有计划的维修措施,这有利于掌握仪器的运行规律,有利于出现故障后的查找。下面
内窥镜设备维修人员来讲讲其保养事项。
The regular maintenance of medical endoscope equipment is generally completed by the instrument controller in cooperation with the engineering technicians. It is an organized and planned maintenance measure that is carried out in a continuous cycle, which is conducive to mastering the operation rules of the instrument and to finding out after the failure. The maintenance personnel of endoscope equipment will talk about its maintenance items.
⑴一保:一般可以一个月一个季度进行一次.主要内容除厂日常维护的工作外,可以拆开机壳,各处积尘,污垢、异物、紧固螺丝、添加润滑剂;检查务器件、元件有无磨损、变形、烧蚀、击穿、松动,受潮,老化、接地不良等情况;检测各组电源电压及纹波,检查高压部件运行和接触情况等;
(1) First maintenance: generally, it can be carried out once a month to once a quarter. In addition to the daily maintenance work of the factory, the main contents can be disassembled, dust, dirt, foreign matter, fastening screws, and lubricant can be removed everywhere; Check the service devices and components for wear, deformation, ablation, breakdown, looseness, moisture, aging, poor grounding, etc; Check the power supply voltage and ripple of each group, and check the operation and contact of high-voltage components;
⑵二保:一般可以半年一年进行一次。主要内容除做好一保外,可以对整机控制台上的各个仪表及操作控制系统的灵敏度,精度进行测试校正和计量检定,更换高压发生器绝缘汕等到期的损耗品,对电路中各测试点的电压,波形进行系统检测和做拉偏试验;
(2) Secondary insurance: generally, it can be carried out once every six months to one year. In addition to the first guarantee, the main contents include testing, calibration and metrological verification of the sensitivity and accuracy of each instrument and operation control system on the console of the whole machine, replacement of expired loss products such as high-voltage generator insulation, systematic detection of the voltage and waveform of each test point in the circuit, and pull bias test;
⑶三保:一般可以2-4年进行一次:主要内容除做好二保外,必要寸可以将整机进行全部拆卸子以清洗检修,超过使用期的元器件应尽量更换或修复,应对仪器进行较为综合的调试,恢复其工作精度和性能,达到或超过新机的程度是完全有可能的。
(3) Three guarantees: generally, it can be carried out once every 2-4 years. In addition to the second guarantee, the whole machine can be fully disassembled for cleaning and maintenance when necessary. Components beyond the service life should be replaced or repaired as much as possible. The instrument should be comprehensively debugged to restore its working accuracy and performance. It is completely possible to reach or exceed the level of the new machine.
The content and time of regular maintenance are different for different instruments. Generally, it can be divided into three levels. The above is the content of the relevant answers to the questions. I hope to give you good help. Please come to our website for more matters http://www.jiangrenyiliao.cn consulting service