喷嘴、水气管堵塞
Nozzle and water and gas pipe are blocked
堵塞原因
Cause of blockage
(1)先端部不小心与硬物碰撞导致喷嘴变形;
(1) The tip accidentally collides with hard objects, resulting in nozzle deformation;
(2)使用内镜后没有立即清洗;
(2) The endoscope was not cleaned immediately after use;
(3)脏物凝固在喷嘴里和水气管道里;
(3) Dirt solidifies in nozzles and water and gas pipelines;
(4)消毒前没有用洗涤剂彻底洗净内镜;
(4) The endoscope was not thoroughly cleaned with detergent before disinfection;
(5)消毒液使蛋白质凝固阻塞喷嘴和水气管;
(5) The disinfectant coagulates the protein and blocks the nozzle and water pipe;
(6)擦试镜面时方向错误;
(6) The direction is wrong when wiping the mirror;
(7)棉纱或纸碎塞在喷嘴里;
(7) Cotton yarn or paper scraps are stuffed in the nozzle;
(8)清洗消毒时,使用有问题的清洗附件;
(8) When cleaning and disinfecting, use problematic cleaning accessories;
(9)使脏物随灌洗送到内镜中。
(9) The dirt is sent to the endoscope with irrigation.
故障检修
Troubleshooting
当喉镜水气不通时,是先把喷嘴卸下然后用导丝从操作部那边通水气管。OLYMPUS的水管和气管在弯曲部位置上通过一个三通跟会合成一条管。
When the water and air of the laryngoscope are blocked, first remove the nozzle, and then use the guide wire to connect the water pipe from the operation Department. The water pipe and gas pipe of Olympus will be combined into a pipe through a tee at the bending position.
维修步骤:
Maintenance steps:
(1)用一号小一字螺丝刀拧松喷嘴螺丝(40型以上多有一个鞘),螺丝不用松出来;
(1) Use a small slotted screwdriver to loosen the nozzle screw (more than 40 types have a sheath), and the screw does not need to be loosened;
(2)将喷嘴取出;
(2) Take out the nozzle;
(3)用导丝在操作部或先端部通水气管,然后加压按钮送水,冲洗水气管道,直到出现很强的一条水流,且送气时气泡很大,然后将水气管中剩余水排出;
(3) Use the guide wire to open the water pipe at the operation part or the tip, then press the button to send water, and flush the water and gas pipe until there is a strong water flow, and the bubbles are large during air supply, and then discharge the remaining water in the water and gas pipe;
(4)用镊子夹镜头纸舔酒精,清洁管道口,直到无污垢;
(4) Use tweezers to clamp lens paper and lick alcohol, and clean the pipe mouth until there is no dirt;
(5)再次试送水送气,防止有污垢掉管道里;
(5) Try to send water and air again to prevent dirt from falling into the pipe;
(6)再用镊子夹镜头纸舔酒精清洁管道口;
(6) Then use tweezers to clamp lens paper and lick alcohol to clean the pipe mouth;
(7)用镊子夹住喷嘴,用小螺丝刀点小许固定黑胶(KE45B)涂于喷嘴的直管部一圈,小心放进管口内,用镊子背面把喷嘴出口压低至管口边,用手摸喷嘴口感觉不刮手,且喷嘴出口要正对IG盖玻璃,然后拧紧喷嘴固定螺丝,在喷嘴边缘和螺丝孔处补胶;
(7) Clamp the nozzle with tweezers, use a small screwdriver to point a small amount of fixed black glue (ke45b) and apply it to the straight pipe of the nozzle for a circle, carefully put it into the nozzle, use the back of the tweezers to press the nozzle outlet down to the edge of the nozzle, touch the nozzle mouth with your hand to feel that you don't scratch your hands, and the nozzle outlet should be facing the Ig cover glass, then tighten the nozzle fixing screws, and fill the glue at the edge of the nozzle and the screw hole;
(8)试送水、送气,要求水流是一条直线冲洒镜面,且水流覆盖镜面90%,送气时气泡要很大,如没达到要求,重复以上步骤;
(8) Try to send water and air. It is required that the water flow should be a straight line to sprinkle the mirror, and the water flow covers 90% of the mirror. The bubbles should be large when sending air. If the requirements are not met, repeat the above steps;
(9)再次补胶,在白炽灯下烘15分钟,再次检查送水送气效果。
(9) Replenish the glue again, bake it under the incandescent lamp for 15 minutes, and check the effect of water and air supply again.
怎样避免喷嘴、水气管堵塞
How to avoid nozzle and water pipe blockage
(1)操作内镜要小心;
(1) Be careful when operating the endoscope;
(2)内镜使用后应立即清洗;
(2) Endoscopes should be cleaned immediately after use;
(3)内镜消毒前必须彻底洗净;
(3) Endoscope must be thoroughly cleaned before disinfection;
(4)擦拭镜面时应顺着喷嘴的方向;
(4) Wipe the mirror along the direction of the nozzle;
(5) The endoscope equipment maintenance personnel believe that before cleaning the endoscope, they should thoroughly check whether the cleaning accessories are defective.
The above is the relevant content of the question answer. I hope it can help you. If you still have any questions about this problem, you are welcome to follow our website http://www.jiangrenyiliao.cn And consult our staff, will serve you wholeheartedly.