随着内窥镜诊疗技术的发展,软式内窥镜广泛运用在肿瘤的筛查、疾病的诊断和,它使用后清洗是所有使用者所担心的问题,下面
内窥镜设备维修人员就来讲下它的清洗处理流程吧。
With the development of endoscopic diagnosis and treatment technology, flexible endoscopes are widely used in tumor screening, disease diagnosis and treatment. Cleaning and disinfection after use is a concern of all users. Let's talk about the cleaning and disinfection process of endoscope equipment maintenance personnel.
一、床旁预处理流程:当内窥镜从病人体内取出后,护士及时使用含有清洗剂的纱布沿着操作部向先端部方向擦拭。然后使用含酶清洗剂的水反复吸引冲洗管腔。
1、 Bedside pretreatment process: after the endoscope is taken out of the patient, the nurse uses gauze containing cleaning agent to wipe along the operation part toward the tip in time. Then, the lumen was repeatedly sucked and rinsed with water containing enzyme detergent.
二、运送室,室工作人员清洗槽内注入适量过滤水(符合内窥镜清洗规范)进行内窥镜的测漏检查,内窥镜完好无渗漏。含清洗液(酶液)的清洗槽内,及时对内窥镜进行预清洗。
2、 After being transported to the disinfection room, the staff of the disinfection room injected an appropriate amount of filtered water into the cleaning tank (in accordance with the endoscope cleaning and disinfection specifications) for leak detection inspection of the endoscope. The endoscope was in good condition without leakage. In the cleaning tank containing cleaning solution (enzyme solution), timely pre clean the endoscope.
1、使用清洁纱布在清洗槽内(含酶清洗剂)反复擦洗镜身,擦洗插入部和操作部,纱布一用一更换。
1. Use cleaning gauze to repeatedly scrub the lens body in the cleaning tank (including enzyme detergent), scrub the insertion part and the operation part, and replace the gauze once in use.
2、酶液下手工刷洗,使用专用内窥镜清洗刷,分别对内窥镜的所有管道、注水注气按钮、先端帽等进行反复刷洗。尤其是按钮内部的空隙,更是刷洗的部位。清洗刷一用一换,清洗液一用一换。
2. Manual brushing under enzyme solution: use special endoscope cleaning brush to repeatedly brush all the pipes, water and gas injection buttons and tip caps of the endoscope. In particular, the gap inside the button is the part to be brushed. The cleaning brush shall be replaced once in use and the cleaning liquid shall be replaced once in use.
3、使用过滤水冲洗干净内窥镜的内管道和外管壁的清洗剂,压力气枪吹干水分。按钮和阀门、清洗刷冲洗干净。
3. Rinse the cleaning agent of the inner pipe and outer pipe wall of the endoscope with filtered water, and blow the water dry with a pressure air gun. The button, valve and cleaning brush shall be cleaned.
4、注气注水按钮、清洗刷、可以复用的附件等放入超声清洗器中用含酶清洗液震荡清洗。
4. Put the gas injection and water injection button, cleaning brush, reusable accessories, etc. into the ultrasonic cleaner and shake them with the cleaning liquid containing enzymes.
三、将过预清洗后的内窥镜及附件、清洗刷共同放入全自动内窥镜清洗机中进行清洗。
3、 Put the pre cleaned endoscope, accessories and cleaning brush into the full-automatic endoscope cleaning and disinfection machine for cleaning.
四、清洗机清洗程序运行结束后,人员更换手套,取出内窥镜并刷卡记录,使用压力气枪向所有管道充气少30s,使其管腔完全干燥。使用无菌擦拭巾、压力气枪干燥内窥镜外表面、按钮和阀门。安装按钮和阀门。将内窥镜镜头涂抹保护蜡。
4、 After the cleaning procedure of the cleaning machine is completed, the disinfection personnel shall replace gloves, take out the endoscope and swipe the card for recording, and inflate all pipes with pressure air gun for at least 30s to completely dry their lumens. Dry the external surface of the endoscope, buttons and valves with sterile wipes and pressure air guns. Install buttons and valves. Apply protective wax to the endoscope lens.
五、放入内窥镜专用运送车备用,不再使用的内窥镜放入专用柜储存,每次使用前进行再次。
5、 Put it into the special delivery vehicle for endoscopes for standby. Put the endoscopes that are no longer in use into the special cabinet for storage. Disinfect them again before each use.
六、对清洗设备及使用的内窥镜每季度会定期进行生物学培养监测清洗质量,使用内窥镜追溯系统,做到每一条内窥镜从诊疗到每一个环节都有据可查。
6、 The cleaning and disinfection equipment and endoscopes used will be regularly biological cultured every quarter to monitor the cleaning and disinfection quality, and the endoscope tracing system will be used to ensure that every link from diagnosis and treatment to disinfection of each endoscope is documented.
It is the user's responsibility to clean and disinfect each endoscope. For more information about the use of endoscopes, please visit our website http://www.jiangrenyiliao.cn Consult and understand!