内镜光电耦合器(CCD)损坏怎么办?

来源:http://www.jiangrenyiliao.cn/ 发布时间:2023-01-29 浏览量:0

电子内镜先端部CCD是内镜的心脏部分,作为微型摄像系统,其对水汽非常敏感。内镜诊疗、洗消都是在液体环境下进行的,任何环节,外表皮、内管道一旦破损,失去密封性,液体进入易导致CCD损坏。内镜光电耦合器(CCD)损坏怎么办?内窥镜维修公司为您提出建议。
The CCD at the tip of the electronic endoscope is the heart of the endoscope. As a miniature camera system, it is very sensitive to water vapor. Endoscopic diagnosis, treatment and decontamination are carried out in a liquid environment. In any link, once the outer skin and inner pipe are damaged, the sealing will be lost, and the liquid will easily cause CCD damage. What if the endoscope optocoupler (CCD) is damaged? The endoscope maintenance company provides suggestions for you.
插入部前端约10cm包裹的橡皮很薄,因此避免碰到坚硬的物体和器械,为减少磨损,进镜、退镜时避免内镜与口垫摩擦。内镜先端部有各种精密结构,因此在取异物尤其是锋利异物时,避免异物过分紧贴镜面,防止刺破、划伤内镜。
The rubber wrapped about 10cm in the front of the insertion part is very thin, so avoid touching hard objects and instruments. In order to reduce wear, avoid friction between the endoscope and the mouth pad when entering and exiting the mirror. There are various precision structures at the tip of the endoscope. Therefore, when removing foreign objects, especially sharp foreign objects, the foreign objects should not be too close to the mirror surface to prevent puncturing and scratching the endoscope.
内镜镜身为合成树脂保护层,形成内部与外界隔绝的密闭环境,一旦漏气,分泌物、黏液、水等进入内窥镜内部,腐蚀其结构[6]。所以为了及时发现内镜损坏,每天检查结束后及时测漏,发现漏气及时将内镜取出吹干并维修处理。
The endoscope body is a synthetic resin protective layer, forming a closed environment that is isolated from the outside. In case of air leakage, secretions, mucus, water, etc. enter the inside of the endoscope and corrode its structure [6]. Therefore, in order to find the damage of the endoscope in time, the leakage shall be detected in time after the inspection every day, and the endoscope shall be taken out and dried in time and repaired in case of air leakage.
内窥镜维修公司
为确保内镜诊疗的顺利进行,操作前要认真检查。送气送水和吸引确保通畅,大小旋钮处于自由状态,光线亮度调节适宜。图像清晰,乙醇纱布顺着喷嘴方向擦拭镜面,必要时白平衡。送气送水按钮、吸引按钮在消毒干燥后涂少许硅油再按在内镜上,活检钳的关节处涂医用硅油。从而使按钮使用时更灵活,避免按钮按下去之后无法弹出。
To ensure the smooth progress of endoscopic diagnosis and treatment, carefully check before operation. The air supply, water supply and suction are smooth, the size knob is in a free state, and the light brightness is adjusted properly. The image is clear, the ethanol gauze wipes the mirror surface along the nozzle direction, and white balance is needed. After disinfection and drying, apply a little silicone oil on the air and water supply buttons and suction buttons, and then press them on the endoscope. Apply medical silicone oil on the joint of the biopsy forceps. This makes the button more flexible to use and prevents the button from being ejected after being pressed down.
内镜价格昂贵,并且相对容易被损坏。因此工作人员不仅要有专业的操作技术,同时要加强责任心教育,定期组织学习培训,做到能够及时处理常见故障,对内镜定期保养及日常维护,及时发现并处理问题,使内镜保持良好工作状态。
Endoscopy is expensive and relatively easy to be damaged. Therefore, the staff should not only have professional operating skills, but also strengthen the sense of responsibility education, regularly organize learning and training, so as to be able to deal with common faults in time, regularly maintain and daily maintain the endoscope, find and deal with problems in time, and keep the endoscope in good working condition.
只有做到这些,才能达到减少维修费用,为科室减少损失,同时达到为患者提供高质量医疗服务的目的。更多相关维修事项就来我们网站http://www.jiangrenyiliao.cn咨询吧!
Only by doing this can we achieve the goal of reducing maintenance costs, reducing losses for departments and providing high-quality medical services for patients. For more relevant maintenance matters, please visit our website http://www.jiangrenyiliao.cn Ask!
s
新闻资讯
联系我们
  • 公司公司:匠仁医疗设备有限公司
  • 匠仁医疗山东济南维修部
  • 联系人:樊经理
  • 电话:13153199508
  • 地址:山东省济南市槐荫区美里东路3000号德迈国际信息产业园6号楼101-2室
  • 匠仁医疗湖南长沙维修部
  • 联系人:李经理
  • 电话:13873135765
  • 地址:湖南省长沙市雨花区劳动东路820号恒大绿洲14栋2409室
新闻资讯

樊经理:13153199508     李经理:13873135765

公司地址:山东省济南市槐荫区美里东路3000号德迈国际信息产业园6号楼101-2室  湖南省长沙市雨花区劳动东路820号恒大绿洲14栋2409室

     

鲁公网安备 37010402441281号