内窥镜设备结构精良、操作自如、图像清晰、使用方便等优点,其精密复杂和娇贵的程度是可想而知的。因此,正确的操作过程和良好的维修保养可延长其使用寿命,常见故障的预防及处理措施也尤为重要。下面
内窥镜设备维修人员为您解答。
Endoscope equipment has the advantages of excellent structure, easy operation, clear image and convenient use, and its precision, complexity and delicate degree can be imagined. Therefore, the correct operation process and good maintenance can extend its service life, and the prevention and treatment measures of common faults are particularly important. The endoscope equipment maintenance personnel will answer for you.
弯曲橡皮
Bending rubber
在镜子体插入部前端有一长约10cm的可以自如弯曲的部分,它是由很薄的橡皮包裹着的,是整个镜体“娇气”的部分,如不慎将其碰到台车或桌面、清洗消毒槽的边缘,或尖硬的物体以及器械等,都会导致弯曲橡皮出现破损或小孔而使镜体失去密封性。因此,操作者在检查完后都应十分小心谨慎。在清洗消毒的过程中应特别注意避免与尖硬、锐利的器械一同清洗消毒。
At the front end of the insertion part of the mirror body, there is a flexible part about 10 cm long, which is wrapped by a very thin rubber and is the "delicate" part of the whole mirror body. If it accidentally touches the edge of the trolley or table top, the cleaning and disinfection tank, or sharp and hard objects and instruments, it will lead to damage or small holes in the bending rubber and make the mirror body lose its sealing. Therefore, operators should be very careful after inspection. In the process of cleaning and disinfection, special attention should be paid to avoid cleaning and disinfection with sharp and hard instruments.
钳道管
Clamp tube
在内窥镜设备检查和治疗过程中,钳道管是检查和治疗器械的必经之道,如在送器械的过程中操作不当,很容易导致钳道管破损而失去镜体的密封性造成漏液,损坏摄像系统。在从钳道管送器械的过程中应注意以下几个方面:
During the inspection and treatment of the endoscope equipment, the clamp tube is the only way to inspect and treat the instrument. If the instrument is delivered improperly, it is easy to cause the clamp tube to be damaged and lose the sealing of the lens body, causing fluid leakage and damaging the camera system. Pay attention to the following aspects in the process of delivering instruments from the forceps:
①先将镜身放直,送器械的动作要轻,用力适量,尽量减少对钳道管内壁的磨损,待器械从钳道管伸出,再调节角度。
① First, straighten the mirror body, and send the instrument gently with appropriate force to minimize the wear on the inner wall of the clamp tube. After the instrument extends from the clamp tube, adjust the angle.
②做电凝或激光治疗时,一定要在探头伸出钳道管并对准治疗部位时再踏开关。
② When performing electrocoagulation or laser treatment, be sure to step on the switch when the probe extends out of the clamp tube and aims at the treatment site.
③一旦送器械时遇到阻力,千万不要强行操作。
③ Once you encounter resistance when delivering the instrument, do not force it.
The inspection, treatment, cleaning and disinfection of endoscope equipment are carried out in a liquid environment. Therefore, the sealing of the whole lens body is particularly important. The common parts that cause the loss of sealing of the lens body are bent rubber and clamp tube. These are the answers to the questions. If you have any questions, please come to our website http://www.jiangrenyiliao.cn Ask!