内窥镜维修之水洗、或者事项

来源:http://www.jiangrenyiliao.cn/ 发布时间:2022-08-11 浏览量:0

内窥镜属于精密仪器,在日常使用过程中经常会有磨损,损坏,所以维修不可避免,那么内窥镜维修电话小编下面就来讲讲水洗、或者事项吧。
Endoscope is a precision instrument, which is often worn and damaged during daily use. Therefore, maintenance is inevitable. Let's talk about washing, disinfection or sterilization in the following.
水洗
washing
1、将内窥镜放入水洗槽,在流动水下冲洗,用纱布擦洗镜身及操作部。纱布一用一换。
1. Put the endoscope into the water washing tank, wash it under flowing water, and scrub the endoscope body and operating part with gauze. The gauze is used and changed.
2、取下活检入口阀门、吸引器按钮和送气送水按钮,用清洁毛刷刷洗活检孔道和导光软管的吸引器管道少3次,刷洗时应两头见刷手,并清洗刷头上的污物。清洗刷一用一。取下的各类阀门、按钮用清水冲洗干净并擦干。
2. Take down the biopsy inlet valve, aspirator button and air and water supply button, and brush the aspirator pipe of the biopsy duct and light guide hose with a cleaning brush for at least 3 times. When brushing, the brush hands should be seen at both ends, and the dirt on the brush head should be cleaned. One for use and one for disinfection. The removed valves and buttons shall be washed with clean water and wiped dry.
3、安装全管道灌流器、管道插塞、防水帽和吸引器,用吸引器反复抽吸活检孔道。
3. Install the whole pipe perfusion device, pipe plug, waterproof cap and aspirator, and use the aspirator to repeatedly aspirate the biopsy channel.
4、全管道灌流器接50ml注射器,吸清水注入送水管道。
4. The whole pipe irrigator is connected with a 50ml syringe, and sucking water is injected into the water supply pipe.
5、用吸引器吸干活检孔道的水分并擦干镜身。
5. Use a suction device to absorb the water in the biopsy channel and dry the lens body.
6、内窥镜附件如活检钳、细胞刷、切开刀、导丝、碎石器、网篮、造影导管、异物钳等使用后,先放入清水中,用小刷刷洗钳瓣内面和轴节处,清洗后擦干。
6. After the endoscope accessories such as biopsy forceps, cell brushes, cutting knives, guide wires, lithotrippers, net baskets, contrast catheters, foreign body forceps, etc. are used, first put them into clean water, use a small brush to wash the inner surface of the forceps and the shaft shaft shaft, and then wipe them dry.
内窥镜维修电话
Disinfection or sterilization
1、采用剂浸泡或者时,应将清洗擦干后的内窥镜置于槽并全部浸没于液中,各孔道用注射器灌满液。
1. When using disinfectant for immersion disinfection or sterilization, the cleaned and dried endoscope shall be placed in the disinfection tank and completely immersed in the disinfectant. Each hole shall be filled with disinfectant with syringe.
2、非全浸式内窥镜的操作部,应用清水擦拭后再手75%乙醇擦拭。
2. The operation part of the non full immersion endoscope shall be wiped with clean water and then wiped with 75% ethanol.
3、时间参见卫生部《内窥镜清洗技术操作规范》或剂、器械的产品使用说明。
3. For the disinfection and sterilization time, please refer to the technical operation specification for endoscope cleaning and disinfection of the Ministry of health or the product instructions for disinfectants and disinfection instruments.
4、内窥镜附件应一用一,选压力蒸汽方法,也可用环氧乙烷或者符合卫生行政部门相关规定的其他方法。
4. The endoscope accessories shall be sterilized with one for use, and the pressure steam sterilization method shall be selected. Ethylene oxide or other sterilization methods conforming to the relevant provisions of the health administrative department may also be used.
5、相关物品清洗后,弯盘、敷料缸等用压力蒸汽;复用的口圈、注水瓶及连接管使用剂浸泡,复用的口圈再用流动水冲净;注水瓶及连接管用无菌水冲净,干燥备用。使用时注水瓶内应注入无菌水,每日更换。
5. After the relevant articles are cleaned, the bending plate and dressing cylinder are sterilized with pressure steam; The reused mouth ring, water injection bottle and connecting pipe shall be soaked and disinfected with disinfectant, and the reused mouth ring shall be flushed with flowing water; Rinse the water injection bottle and connecting pipe with sterile water and dry them for standby. During use, sterile water shall be injected into the water injection bottle and replaced daily.
6、使用机进行清洗之前,应先按照预处理、测漏、水洗、酶洗、清洗的要求对内窥镜进行清洗。
6. Before using the disinfection machine for cleaning and disinfection, the endoscope shall be cleaned according to the requirements of pretreatment, leak detection, water washing, enzyme washing and cleaning.
此外要注意,虽然各家的内窥镜结构和原理基本差不多,但是由于选用的配件规格型号不一样,往往零件也并不能够通用,所以基本上是不能够找到其它代修的。更多注意事项就来我们网站http://www.jiangrenyiliao.cn咨询吧。
In addition, it should be noted that although the structure and principle of endoscopes of various brands are basically the same, because the specifications and models of the selected accessories are different, the parts are often not universal, so it is basically impossible to find other brands to repair. For more information, please visit our website http://www.jiangrenyiliao.cn Consult.
s
新闻资讯
联系我们
  • 公司公司:匠仁医疗设备有限公司